Hioki TEMP UNIT 8967 is a high-precision temperature measurement module designed for energy storage and battery testing applications. Supporting temperature waveform measurement from -200°C to 2000°C, it ensures accurate data acquisition for critical thermal analysis. Compatible with Hioki Memory HiCorders like MR6000, MR8847A, and MR8827, this module offers 2 channels with thermocouple input support (K, J, E, T, N, R, S, B, W), delivering high-resolution temperature monitoring. Featuring a wide measurement range, fast data refresh rates, and internal/external reference junction compensation, it is ideal for battery R&D, energy storage system validation, and industrial thermal testing in demanding environments.More
  • TEMP UNIT 8967 High-Precision Temperature Measurement Module for Energy Storage & Battery Testing Hioki
  • TEMP UNIT 8967 High-Precision Temperature Measurement Module for Energy Storage & Battery Testing Hioki
8967For MR6000, MR8847A, MR8827, and similar products
Note: This module must be used with the Memory HiCorder. Input cords are not included. Please purchase them separately.

Basic specifications

Accuracy guaranteed: 1 year
Measurement functionsNumber of channels: 2, for temperature measurement with thermocouple (voltage measurement not available)
Input connectorsThermocouple input: plug-in connector,Recommended wire diameter: single-wire, 0.14 to 1.5 mm2, braided wire 0.14 to 1.0 mm2 (conductor wire diameter min. 0.18 mm), AWG 26 to 16Input impedance: min. 5 MΩ (with line fault detection ON/OFF),Max. rated voltage to earth: 300 V AC, DC(with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage)
Temperature measurement rangeNote: Upper and lower limit values depend on the thermocouple10 °C/div (-100 °C to 200 °C), 50 °C/div (-200 °C to 1000 °C), 100 °C/div (-200 °C to 2000 °C), 3 ranges, full scale: 20 div,Measurement resolution: 1/1000 of measurement range (using 16-bit A/D conversion and when installed in the 8847)
Thermocouple range(JIS C 1602-1995)(ASTM E-988-96)K: -200 to 1350 °C, J: -200 to 1100 °C, E: -200 to 800 °C, T: -200 to 400 °C, N: -200 to 1300 °C, R: 0 to 1700 °C, S: 0 to 1700 °C,B: 400 to 1800 °C, W (WRe5-26): 0 to 2000 °C,Reference junction compensation: internal/ external (switchable),Line fault detection ON/OFF possible
Data refresh rate3 methods, Fast: 1.2 ms (digital filter OFF), Normal: 100 ms (digital filter 50/60 Hz), Slow: 500 ms (digital filter 10Hz)
Measurement accuracyThermocouple K, J, E, T, N: ±0.1 % of full scale ±1 °C (±0.1 % of full scale ±2 °C at -200 °C to 0 °C),Thermocouple R, S, W: ±0.1 % of full scale ±3.5 °C (at 0 °C to 400 °C or less), ±0.1 % of full scale ±3 °C (at 400 °C or more)Thermocouple B: ±0.1 % of full scale ±3 °C (at 400 °C or more),Reference junction compensation accuracy: ±1.5 °C (added to measurement accuracy with internal reference junction compensation)
Dimensions and mass106 mm (4.17 in) W × 19.8 mm (0.78 in) H × 204.5 mm (8.05 in) D, 240 g (8.5 oz)
Included accessoriesFerrite clamp × 2
Download
REQUEST MORE DETAILS
Please fill out the form below and click the button to request more information about
Fill out the form below to make an inquiry
Company*
Country/Region*
FirstName*
LastName*
Email*
Whatsapp/Phone*
Product Description*
Verification code*
We needs the contact information you provide to us to contact you about our products and services.
If your supplier does not respond within 24 hours, we will connect you with three to five qualified alternative suppliers.
我們使用 Cookie 來改善您的線上體驗。 繼續瀏覽本網站,即表示您同意我們使用 Cookie